Plate Material Evaluating Cell
This quite handy cell is developed in order to evaluate plate material such as metal, semi-conducting plate, etc.
In evaluation, a sample plate is sandwiched between cell blocks.
Catalog No. | Description | Quantity |
011951 | Plate Material Evaluating Cell | 1 |
Contents | ||
Teflon Cell [Body] | 1 | |
Teflon Cell [Base] | 1 | |
O-ring | 1 | |
Screw 20 mm | 2 | |
Purging tube | 1 | |
002222 | Pt counter electrode for VC-2 cell | 1 |
[Optional Products] Reference Electrode
Users can select a reference electrode separately for this cell from :
Catalog No. | Description |
012167 | RE-1B Reference electrode (Ag/AgCl) |
012171 | RE-7 Non Aqueous reference electrode (Ag/Ag+) |
002058 | RE-1C Reference electrode (Ag/AgCl/Saturated KCl) |
002056 | RE-2B Calomel Reference electrode |
002057 | RE-2C Reference electrode |
012974 | RE-6A Reference electrode for alkaline solution |
How to Set Up
1. Place a plate sample on a lower side teflon block of the cell.
Applicable plate dimensions : W: 10-20 mm, T: ~3 mm
2. Put another teflon block on the sample, and drive two screws evenly to fix the plate. Liquid contact area depends on inner diameter of the O-ring. Deliver approximately 1 ml of electrolyte after fixing the plate.
3. Attach a Pt counter electrode and a reference electrode to the cell by setting them onto a cell cap. We would recommend you to twist up the Pt part appropriately if the counter electrode is longer.